 |  Anthropomorphic
 Molded Around Skull
 Checks Important Physical Parameters
 Provides Realistic System Check
 Ideal For Training
 Polycarbonate Base Dimensioned For Scans Parallel or Perpendicular to Longitudinal Axis SECTION A: Aluminum plates, 0.4mm thick, set at a 45° angle, measure beam width and slice thickness.
 
 SECTION B: The dosimetry section provides patient-exposure controls, and checks on abnormal internal doses before
actual patient scans. Customization to accommodate TLD with rods or chips is available.
 
 SECTION C: Cylindrical tumors of graded sizes and radiodensities establish high and low-contrast resolution and
demonstrate partial beam-averaging effects.
 
 SECTION D: The anthropomorphic section provides actual auditory ossicles in an air-contrast inner ear and a
fractured petrous bone. Small tumors in the posterior fossa are placed in an area which most scanners image with
difficulty.
 
 SECTION E: 1/4 Inch diameter aluminum rod on the phantom's longitudinal axis, checks general alignment, and
"DryWater" gives a realistic "noise check."
 
 
 
 |