|
Main Module: Set of PMMA attenuation plates: 6x10mm thick and 2x5mm thick;
one plate contains marking of the reference point.
|
|
CNR Module: 10mm thick module containing a 20 x 20 x 0.2mm aluminium filter,
located 6 cm from the chest side.
|
|
Contrast Detail Module: 20mm thick module containing gold (99.99%) discs organized in a
7x14 matrix (diameter x thickness). Discs have following diameters: 0.1, 0.25, 0.5,
0.75, 1.0, 1.5, 2.0mm and 14 thicknesses ranging from 0.03 to 2.0µm. Thickness
accuracy: 1nm (0.001 µm), diameter accuracy 0.001mm (1µm)
|
|
Ghosting Module: 10mm thick module containing a 30 x 30 x 0.1mm aluminum filter for a
ghost test.
|
|
Resolution and Geometry Module: 10mm thick module containing:
pattern for the line pair resolution evaluation (from 1,5 to 20,0 LP/mm) rotated 45°
8 low contrast objects (ø 5.5mm and depth from 0.1mm to 0.45mm)
3 objects of a different absorption level.
pattern for evaluation of the effective radiation field.
|
|
MTF Module: 10mm thick module containing a straight stainless steel edge accurate
to ±2µm at a 3° angle.
|
|
Artifact Evaluation Module: 10mm thick module containing a mesh for artifact evaluation.
|
|
Filaments Module: 10mm thick module containing 6 groups of multi-directional filaments
0.40mm to 0.20mm in diameter.
|
|
Dynamic Range Module: 10mm thick module containing Al step wedge with 14 steps from 0.0 to 5.2mm
|
|
Geometry Distortion Module (scales): 10mm thick module containing a grid with scales.
|
|
Geometric Distortion Module (mesh): 10mm thick module containing a wire mesh of horizontal,
vertical and diagonal lines (45°)
|
|
ACR Accreditation Module: 14mm thick module containing a wax insert like in Pro-MAM
Accreditation (when used with 3x10mm plates from main module simulates 42 mm compressed
breast of average glandular/adipose composition.)
|
|
Full Field ACR Accreditation Module: 14mm thick module containing a wax insert like
in Pro-MAM Accreditation FF (when used with 3x10mm plates from main module simulates 42 mm
compressed breast of average glandular/adipose composition.)
|
|
Noise Evaluation Module: 2mm thick aluminum plate 200x200mm.
|
|
Spacers Sets: 180x15 or 240x15mm PMMA plates:
4 pieces 10 mm thick,
2 pieces 8 mm thick,
2 pieces 5 mm thick,
2 pieces 2 mm thick.
|
|
Compensation Module: set of 10 PMMA plates 40x20x2mm
|