| Source-Ray DIRECT-DIGITAL Portable X-Ray Systems. | 
| Stand Weight: 100 lbs. / 45 Kg PowerMax 1260 X-Ray Generator System Features... 
  The PowerMax 1260 provides the maximum power obtainable from a line powered portable with state
  of the art technology, system reliability and low maintenance, all in a small footprint, lightweight
  package. The PowerMax 1260 is a 120kV, 15-60mA system. With 3kW across the full range,
  it can reduce exposure times on difficult radiographic anatomical views, thereby minimizing patient
  motion problems and enhancing image quality.
   
   kVp Range: 40 - 120 kVp, adjustable in 1kV increments with closed loop regulation mA Range: Adjustable 15 - 60 mA, with closed loop regulation mAs Range: 0.125 - 200 Total Max Power Output: 3kW Input Voltage: 110 - 230VAC 50/60Hz (Power Factor Correction) Exposure Time: 0.01 - 3.33 seconds Indicators: Digital display of kVp, mAs and Exposure Time, X-Ray, Ready, mA Station and Fault indication. Beam Filtration: 3.6 mm Al minimum @ 100kVp, 4.3 mm al minimum @ 120kVp Exposure Switch: Detachable two-position hand switch X-Ray Tube... 
    Type: Stationary Anode, 130 kVpCollimator... Target Material: Tungsten Anode Heat Storage Capacity: 30,000 HU Focal Spot: 1.2mm Target Angle: 15° 
    Make/Model: Collimare CP-1 00-241-CTravel Stand... Laser Crosshairs: Center Locator Tape Measure: SID Indicator Inclinometer: Angle Indicator. 
    Min. Focal Spot to Floor: 13"/33 cm Max. Focal Spot to Floor: 78" /198 cm Horiz. Ext. of F/S: 21"/53 cm Horiz. Rotation: 200° Tube Head Rotation: 270° Stand Weight: 110 lbs. / 50 Kg |  | 
| 
 | 
|  | 
 |