Ludlum Model 375 Series Radiation Area Monitors. |
FEATURES: INTRODUCTION:
The Model 375 is a versatile, compact and very affordable digital electronic controller designed for monitoring
radiation in areas. Its simple design accommodates many different detectors suiting a wide variety of applications and
is equipped with a local readout and alarms. These versatile units may also be connected to an optional auxiliary
indicator/annunciators for alerting personnel at remote locations. The user-friendly, digital design enhances setup and
operation. These units may also be networked to a central PC-based station where data are logged and alarms posted.
SPECIFICATIONS:
INDICATED USE: Area monitoring
SUGGESTED DETECTORS: GM, proportional, scintillation, neutron DISPLAY: 4 digit LED display with 0.8" (2 cm) digits DISPLAY RANGE: 000.0 - 9999 DISPLAY UNITS: Can be made to display in R/hr, mR/hr, R/hr, Sv/hr, mSv/hr, Sv/hr, rem/hr, mrem/hr, rem/hr, cpm, cps, and others LINEARITY: Reading within 10% of true value RESPONSE: Typically 3 seconds from 10% - 90% of final reading STATUS: Green light, instrument functioning properly ALARMS:
A. Low Alarm: Yellow light, 1 beep/second audible, selectable range: 0-9999
HIGH VOLTAGE: Adjustable from 200 - 2500 voltsB. High Alarm: Red light, 4 beeps/second audible, selectable range: 0-9999 C. Detector Fail: Red light, constant audible tome > 68 dB at 2 feet D. Low Bat: Yellow light, indicates less than 2 hours of battery life remaining THRESHOLD: Adjustable from 2 - 100 mV DEAD TIME: Adjustable to compensate for dead time of detector and electronics (can be read on display) OVERLOAD: Senses detector saturation (indicated by display reading "-OL-") OVERRANGE: Indicates radiation field being measured has exceeded counting range of instrument (indicated by display reading) DATA OUTPUT: 9 pin connector providing 5 decade logarithmic output, RS-232 output, signal ground connection, FAIL and ALARM signals (current sink), and direct connection to battery and ground. CALIBRATION CONTROLS: Accessible from front of instrument (protective cover provided). POWER: 95 - 135 Vac (178 - 240 Vac available), 50-60 Hz single phase. 6 volt sealed lead-acid rechargeable battery (built in) BATTERY LIFE: Typically 48 hours in non-alarm condition; 12 hours in alarm condition. BATTERY CHARGER: Battery is continuously trickle charged when instrument is connected to line power and turned on. CONSTRUCTION: Wall mount aluminum housing with ivory powder coat paint. SIZE: 7.4" H x 9.7" W x 2.5" D (18.7 x 24.6 x 6.4 cm) WEIGHT: 6.5 lbs. (2.3 kg) OPTIONS DESCRIPTION:
ETHERNET OPTION: Enables the Model 375 Digital Controller to be connected to an Ethernet Network. If ordering
this option, please download and complete
the Model 375 Network Setting Request Form and submit it with your order so the instrument can be properly
configured at Ludlum's factory to facilitate your network system. CAMERA OPTION: An industrial camera that delivers high speed IP video, enhanced low-light performance, and 1280 x 1024 resolution. This can be used to capture a picture of what triggers an alarm using the Ethernet Software Application. |
|
Ethernet/Networking Hardware and Software Options. |
|